1Pifzer
3Plum A+3 Infusion System Firmware
Plum A+ Infusion System FirmwareSymbiq Infusion System Firmware
Nov 21, 2024
Mar 25, 2019
N/A· v4
9.8 CRITICAL· v3
10.0 HIGH· v2
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unau...Show more
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.Show less
1Pifzer
3Plum A+3 Infusion System Firmware
Plum A+ Infusion System FirmwareSymbiq Infusion System Firmware
Nov 21, 2024
Mar 25, 2019
N/A· v4
9.8 CRITICAL· v3
10.0 HIGH· v2
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by...Show more
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.Show less
1Pifzer
3Plum A+3 Infusion System Firmware
Plum A+ Infusion System FirmwareSymbiq Infusion System Firmware
Nov 21, 2024
Mar 25, 2019
N/A· v4
9.8 CRITICAL· v3
10.0 HIGH· v2
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends th...Show more
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.Show less
1Pifzer
3Plum A+3 Infusion System Firmware
Plum A+ Infusion System FirmwareSymbiq Infusion System Firmware
Nov 21, 2024
Mar 25, 2019
N/A· v4
7.5 HIGH· v3
5.0 MEDIUM· v2
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends th...Show more
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.Show less