Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Affected (3)
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Up to 13.4 |
| Running on/with | Platform Versions |
|---|---|
Pifzer Plum A+ Infusion System | All versions |
Configuration B
| Vulnerable Software | Affected Versions |
|---|---|
| Up to 13.6 |
| Running on/with | Platform Versions |
|---|---|
Pifzer Plum A+3 Infusion System | All versions |
Configuration C
| Vulnerable Software | Affected Versions |
|---|---|
| Up to 3.13 |
| Running on/with | Platform Versions |
|---|---|
Pifzer Symbiq Infusion System | All versions |
Related CWEs
CWE-200
Exposure of Sensitive Information to an Unauthorized Actor
The product exposes sensitive information to an actor that is not explicitly authorized to have access to that information.
CWE-312
Cleartext Storage of Sensitive Information
The product stores sensitive information in cleartext within a resource that might be accessible to another control sphere.
References (2)
Source: ics-cert@hq.dhs.gov
MitigationThird Party AdvisoryUS Government Resource