CVEs (4)
CVE VENDORS PRODUCTS UPDATED PUBLISHED CVSS |
|---|
1Pifzer 3Plum A+3 Infusion System Firmware Plum A+ Infusion System FirmwareSymbiq Infusion System FirmwareNov 21, 2024 Mar 25, 2019 N/A· v4 9.8 CRITICAL· v3 10.0 HIGH· v2 Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unau...Show more |
1Pifzer 3Plum A+3 Infusion System Firmware Plum A+ Infusion System FirmwareSymbiq Infusion System FirmwareNov 21, 2024 Mar 25, 2019 N/A· v4 9.8 CRITICAL· v3 10.0 HIGH· v2 Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by...Show more |
1Pifzer 3Plum A+3 Infusion System Firmware Plum A+ Infusion System FirmwareSymbiq Infusion System FirmwareNov 21, 2024 Mar 25, 2019 N/A· v4 9.8 CRITICAL· v3 10.0 HIGH· v2 Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends th...Show more |
1Pifzer 3Plum A+3 Infusion System Firmware Plum A+ Infusion System FirmwareSymbiq Infusion System FirmwareNov 21, 2024 Mar 25, 2019 N/A· v4 7.5 HIGH· v3 5.0 MEDIUM· v2 Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends th...Show more |