← Back

CVE-2015-3953

nvd nist
Published: Mar 25, 2019Modified: Nov 21, 2024

JSON object

Loading...
9.8
Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 3.9 / Impact: 5.9
Source: NVD

Description

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected (3)

3 products
Plum A+ Infusion System Firmware
Plum A+3 Infusion System Firmware
Symbiq Infusion System Firmware
Configuration A
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
Up to 13.4
Running on/withPlatform Versions
Pifzer
Plum A+ Infusion System
All versions
Configuration B
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
Up to 13.6
Running on/withPlatform Versions
Pifzer
Plum A+3 Infusion System
All versions
Configuration C
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
Up to 3.13
Running on/withPlatform Versions
Pifzer
Symbiq Infusion System
All versions

References (2)

Source: ics-cert@hq.dhs.gov
MitigationThird Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
MitigationThird Party AdvisoryUS Government Resource

Timeline

No history available yet.