← Back

Mycarelink Smart Model 25000 Firmware

mycarelink_smart_model_25000_firmware

Vendor: Medtronic • 3 CVEs

CVEs (3)

CVE
VENDORS
PRODUCTS
UPDATED
PUBLISHED
CVSS
1Medtronic
1Mycarelink Smart Model 25000 Firmware
Jun 17, 2026
Dec 14, 2020
N/A· v4
8.1 HIGH· v3
9.3 HIGH· v2
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited,...Show more
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.Show less
1Medtronic
1Mycarelink Smart Model 25000 Firmware
Jun 17, 2026
Dec 14, 2020
N/A· v4
9.8 CRITICAL· v3
10.0 HIGH· v2
Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software...Show more
Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the deviceShow less
1Medtronic
1Mycarelink Smart Model 25000 Firmware
Jun 17, 2026
Dec 14, 2020
N/A· v4
8.8 HIGH· v3
5.8 MEDIUM· v2
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable...Show more
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.Show less