CVE-2020-27252
8.1
Vector
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 2.2 / Impact: 5.9
Source: NVD
Description
Medtronic MyCareLink Smart 25000 is
vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
Affected (1)
Products: Medtronic: Mycarelink Smart Model 25000 Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Mycarelink Smart Model 25000 | All versions |
References (3)
Source: ics-cert@hq.dhs.gov
Source: ics-cert@hq.dhs.gov
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource