CVE-2020-25187
9.8
Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 3.9 / Impact: 5.9
Source: NVD
Description
Medtronic MyCareLink Smart 25000 is
vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
Affected (1)
Products: Medtronic: Mycarelink Smart Model 25000 Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Mycarelink Smart Model 25000 | All versions |
Related CWEs
CWE-122
Heap-based Buffer Overflow
A heap overflow condition is a buffer overflow, where the buffer that can be overwritten is allocated in the heap portion of memory, generally meaning that the buffer was allocated using a routine such as malloc().
CWE-787
Out-of-bounds Write
The product writes data past the end, or before the beginning, of the intended buffer.
References (3)
Source: ics-cert@hq.dhs.gov
Source: ics-cert@hq.dhs.gov
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource
Timeline