CVE VENDORS PRODUCTS UPDATED PUBLISHED CVSS |
|---|
The FACSChorus software does not properly assign data access privileges for operating system user accounts. A non-administrative OS account can modify information stored in the local application data folders. |
The FACSChorus software database can be accessed directly with the privileges of the currently logged-in user. A threat actor with physical access could potentially gain credentials, which could be used to alter or destr...Show more |
The FACSChorus software contains sensitive information stored in plaintext. A threat actor could gain hardcoded secrets used by the application, which include tokens and passwords for administrative accounts. |
The FACSChorus workstation does not prevent physical access to its PCI express (PCIe) slots, which could allow a threat actor to insert a PCI card designed for memory capture. A threat actor can then isolate sensitive in...Show more |
The Operating System hosting the FACSChorus application is configured to allow transmission of hashed user credentials upon user action without adequately validating the identity of the requested resource. This is possib...Show more |
There is no BIOS password on the FACSChorus workstation. A threat actor with physical access to the workstation can potentially exploit this vulnerability to access the BIOS configuration and modify the drive boot order...Show more |
The FACSChorus workstation operating system does not restrict what devices can interact with its USB ports. If exploited, a threat actor with physical access to the workstation could gain access to system information and...Show more |
An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an attacker. |
Alaris Systems Manager does not perform input validation during the Device Import Function. |
A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session. |
A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs.
|
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running. |
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
|
The firmware update package for the wireless card is not properly signed and can be modified. |
The Alaris Infusion Central software, versions 1.1 to 1.3.2, may contain a recoverable password after the installation. No patient health data is stored in the database, although some site installations may choose to sto...Show more |
1Bd 7Bodyguard 121 Twins Firmware Bodyguard 323 Colorvision FirmwareBodyguard 999 603 Firmware+4 moreJun 17, 2026 Dec 5, 2022 N/A· v4 5.3 MEDIUM· v3 N/A· v2 The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or di...Show more |
BD Totalys MultiProcessor, versions 1.70 and earlier, contain hardcoded credentials. If exploited, threat actors may be able to access, modify or delete sensitive information, including electronic protected health inform...Show more |
BD Synapsys™, versions 4.20, 4.20 SR1, and 4.30, contain an insufficient session expiration vulnerability. If exploited, threat actors may be able to access, modify or delete sensitive information, including electronic p...Show more |
1Bd 16Pyxis Anesthesia Station Es Firmware Pyxis Ciisafe FirmwarePyxis Logistics Firmware+13 moreJun 17, 2026 Jun 2, 2022 N/A· v4 8.8 HIGH· v3 8.3 HIGH· v2 Specific BD Pyxis™ products were installed with default credentials and may presently still operate with these credentials. There may be scenarios where BD Pyxis™ products are installed with the same default local operat...Show more |
BD Viper LT system, versions 2.0 and later, contains hardcoded credentials. If exploited, threat actors may be able to access, modify or delete sensitive information, including electronic protected health information (eP...Show more |