← Back

CVE-2022-43557

nvd nist
Published: Dec 5, 2022Modified: Jun 17, 2026

JSON object

Loading...
5.3
Vector
CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H
Exploitability: 0.5 / Impact: 4.7
Source: NVD

Description

The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.

Affected (7)

7 products
Bodyguard 999 603 Firmware
Bodyguard Duo 999 903 Firmware
Bodyguard T 999 103 Firmware
Bodyguard 121 Twins Firmware
Configuration A
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Bd
Bodyguard 999 603
All versions
Configuration B
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Bd
Bodyguard Duo 999 903
All versions
Configuration C
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Bd
Bodyguard Epidural 999 683
All versions
Configuration D
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Bd
Bodyguard Pain Manager 999 803
All versions
Configuration E
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Bd
Bodyguard T 999 103
All versions
Configuration F
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Bd
Bodyguard 323 Colorvision
All versions
Configuration G
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Bd
Bodyguard 121 Twins
All versions

Timeline