← Back

Infinity Explorer C700 Firmware

infinity_explorer_c700_firmware

Vendor: Draeger • 3 CVEs

CVEs (3)

CVE
VENDORS
PRODUCTS
UPDATED
PUBLISHED
CVSS
1Draeger
4Delta Xl Firmware
Infinity Delta FirmwareInfinity Explorer C700 Firmware+1 more
Nov 21, 2024
Jan 28, 2019
N/A· v4
6.5 MEDIUM· v3
3.3 LOW· v2
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing...Show more
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.Show less
1Draeger
4Delta Xl Firmware
Infinity Delta FirmwareInfinity Explorer C700 Firmware+1 more
Nov 21, 2024
Jan 28, 2019
N/A· v4
6.5 MEDIUM· v3
3.3 LOW· v2
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending...Show more
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.Show less
1Draeger
4Delta Xl Firmware
Infinity Delta FirmwareInfinity Explorer C700 Firmware+1 more
Nov 21, 2024
Jan 28, 2019
N/A· v4
7.8 HIGH· v3
7.2 HIGH· v2
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the...Show more
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system.Show less