Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.

Affected (4)

4 products
Kappa Firmware
Infinity Explorer C700 Firmware
Delta Xl Firmware
Infinity Delta Firmware
Configuration A
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Draeger
Kappa
All versions
Configuration B
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Draeger
Infinity Explorer C700
All versions
Configuration C
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Draeger
Delta Xl
All versions
Configuration D
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Draeger
Infinity Delta
All versions

References (4)

Source: ics-cert@hq.dhs.gov
Third Party AdvisoryVDB Entry
Source: ics-cert@hq.dhs.gov
Third Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryVDB Entry
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource

Timeline