CVE-2020-16220
4.3
Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L
Exploitability: 2.8 / Impact: 1.4
Source: NVD
Description
In Patient Information Center iX (PICiX) Versions C.02, C.03,
PerformanceBridge Focal Point Version A.01, the product receives input
that is expected to be well-formed (i.e., to comply with a certain
syntax) but it does not validate or incorrectly validates that the input
complies with the syntax, causing the certificate enrollment service to
crash. It does not impact monitoring but prevents new devices from
enrolling.
Affected (4)
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Version b.02 | |
| Version a.01 |
References (4)
Source: ics-cert@hq.dhs.gov
Third Party AdvisoryUS Government Resource
Source: ics-cert@hq.dhs.gov
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Timeline
No history available yet.