← Back

CVE-2017-12725

nvd nist
Published: Feb 15, 2018Modified: Nov 21, 2024

JSON object

Loading...
5.6
Vector
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
Exploitability: 2.2 / Impact: 3.4
Source: NVD

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Affected (3)

Configuration A
3 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
Smiths Medical
Version 1.1
Version 1.5
Version 1.6
Running on/withPlatform Versions
Smiths Medical
Medfusion 4000 Wireless Syringe Infusion Pump
All versions

References (4)

Source: ics-cert@hq.dhs.gov
Third Party AdvisoryVDB Entry
Source: ics-cert@hq.dhs.gov
Third Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryVDB Entry
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource

Timeline

No history available yet.