← Back

Contechealth

contechealth

5 CVEs • 2 products

Products (2)

Click to collapse
Toggle
Cms8000
cms8000

CVEs (5)

CVE
VENDORS
PRODUCTS
UPDATED
PUBLISHED
CVSS
1Contechealth
1Cms8000 Firmware
Nov 21, 2024
Sep 13, 2022
N/A· v4
5.7 MEDIUM· v3
N/A· v2
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attem...Show more
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.Show less
1Contechealth
1Cms8000 Firmware
Nov 21, 2024
Sep 13, 2022
N/A· v4
4.4 MEDIUM· v3
N/A· v2
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse e...Show more
Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.Show less
1Contechealth
1Cms8000 Firmware
Nov 21, 2024
Sep 13, 2022
N/A· v4
7.5 HIGH· v3
N/A· v2
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device t...Show more
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.Show less
1Contechealth
1Cms8000 Firmware
Nov 21, 2024
Sep 13, 2022
N/A· v4
6.1 MEDIUM· v3
N/A· v2
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables...Show more
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parametersShow less
1Contechealth
1Cms8000 Firmware
Nov 21, 2024
Sep 13, 2022
N/A· v4
6.8 MEDIUM· v3
N/A· v2
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to pr...Show more
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.Show less