← Back

Brilliance Ict Firmware

brilliance_ict_firmware

Vendor: Philips • 3 CVEs

CVEs (3)

CVE
VENDORS
PRODUCTS
UPDATED
PUBLISHED
CVSS
1Philips
4 Brilliance Ct Big Bore Firmware
Brilliance Firmware 64Brilliance Ict Firmware+1 more
Jun 17, 2026
May 4, 2018
N/A· v4
8.7 HIGH· v3
6.8 MEDIUM· v2
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2....Show more
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.Show less
1Philips
4 Brilliance Ct Big Bore Firmware
Brilliance Firmware 64Brilliance Ict Firmware+1 more
Jun 17, 2026
May 4, 2018
N/A· v4
7.8 HIGH· v3
7.2 HIGH· v2
Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed creden...Show more
Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system.Show less
1Philips
4 Brilliance Ct Big Bore Firmware
Brilliance Firmware 64Brilliance Ict Firmware+1 more
Jun 17, 2026
May 4, 2018
N/A· v4
8.8 HIGH· v3
7.2 HIGH· v2
Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, o...Show more
Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.Show less