← Back

Mycarelink 24952 Patient Monitor Firmware

mycarelink_24952_patient_monitor_firmware

Vendor: Medtronic • 2 CVEs

CVEs (2)

CVE
VENDORS
PRODUCTS
UPDATED
PUBLISHED
CVSS
1Medtronic
2Mycarelink 24950 Patient Monitor Firmware
Mycarelink 24952 Patient Monitor Firmware
May 19, 2026
Aug 10, 2018
N/A· v4
4.4 MEDIUM· v3
3.8 LOW· v2
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac...Show more
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.Show less
1Medtronic
2Mycarelink 24950 Patient Monitor Firmware
Mycarelink 24952 Patient Monitor Firmware
Jun 22, 2026
Aug 10, 2018
N/A· v4
7.1 HIGH· v3
1.9 LOW· v2
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.