← Back

Miseq Firmware

miseq_firmware

Vendor: Illumina • 2 CVEs

CVEs (2)

CVE
VENDORS
PRODUCTS
UPDATED
PUBLISHED
CVSS
1Illumina
11Iscan Firmware
Iseq 100 FirmwareMiniseq Firmware+8 more
Jun 17, 2026
Apr 28, 2023
N/A· v4
7.5 HIGH· v3
N/A· v2
Instruments with Illumina Universal Copy Service v2.x are vulnerable due to binding to an unrestricted IP address. An unauthenticated malicious actor could use UCS to listen on all IP addresses, including those capable...Show more
Instruments with Illumina Universal Copy Service v2.x are vulnerable due to binding to an unrestricted IP address. An unauthenticated malicious actor could use UCS to listen on all IP addresses, including those capable of accepting remote communications. Show less
1Illumina
11Iscan Firmware
Iseq 100 FirmwareMiniseq Firmware+8 more
Jun 17, 2026
Apr 28, 2023
N/A· v4
9.8 CRITICAL· v3
N/A· v2
Instruments with Illumina Universal Copy Service v1.x and v2.x contain an unnecessary privileges vulnerability. An unauthenticated malicious actor could upload and execute code remotely at the operating system level, whi...Show more
Instruments with Illumina Universal Copy Service v1.x and v2.x contain an unnecessary privileges vulnerability. An unauthenticated malicious actor could upload and execute code remotely at the operating system level, which could allow an attacker to change settings, configurations, software, or access sensitive data on the affected product. Show less