← Back

Endo Surgery Generator Gen11 Firmware

endo-surgery_generator_gen11_firmware

Vendor: Ethicon • 1 CVE

CVEs (1)

CVE
VENDORS
PRODUCTS
UPDATED
PUBLISHED
CVSS
1Ethicon
1Endo Surgery Generator Gen11 Firmware
May 13, 2026
Dec 5, 2017
N/A· v4
4.8 MEDIUM· v3
3.3 LOW· v2
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon En...Show more
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.Show less