← Back

CVE-2025-24836

nvd nist
Published: Feb 13, 2025Modified: Jun 17, 2026Deferred

JSON object

Loading...
6.1
Vector
CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Show more
CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:XShow less
Source: ics-cert@hq.dhs.gov (Secondary)

Description

With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.