CVE-2025-14097
7.2
Vector
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Exploitability: 1.2 / Impact: 5.9
Source: 46b595e9-1acc-41cb-9398-adaf98d37a9b (Secondary)
Description
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution:
Customers should ensure the following:
• The network is secure, and access follows best practices.
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
References (1)
Source: 46b595e9-1acc-41cb-9398-adaf98d37a9b
Timeline
No history available yet.