CVE-2025-0683
8.2
Vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:XShow more
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:XShow less
Source: ics-cert@hq.dhs.gov (Secondary)
Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text
patient data to a hard-coded public IP address when a patient is hooked
up to the monitor. This could lead to a leakage of confidential patient
data to any device with that IP address or an attacker in a
machine-in-the-middle scenario.
References (5)
Source: ics-cert@hq.dhs.gov
Source: ics-cert@hq.dhs.gov
Source: af854a3a-2127-422b-91ae-364da2661108
Source: af854a3a-2127-422b-91ae-364da2661108
Source: af854a3a-2127-422b-91ae-364da2661108
Timeline
No history available yet.