← Back

CVE-2023-0888

nvd nist
Published: Mar 13, 2023Modified: Nov 21, 2024

JSON object

Loading...
7.2
Vector
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Exploitability: 1.2 / Impact: 5.9
Source: NVD

Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Affected (2)

1 product
Configuration A
2 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
Bbraun
Up to 053l000092
Up to 054u000092
Running on/withPlatform Versions
Bbraun
Battery Pack Sp With Wifi
All versions

References (4)

Source: productsecurity@bbraun.com
Product
Source: productsecurity@bbraun.com
Product
Source: af854a3a-2127-422b-91ae-364da2661108
Product
Source: af854a3a-2127-422b-91ae-364da2661108
Product

Timeline