← Back

CVE-2021-37165

nvd nist
Published: Aug 2, 2021Modified: Nov 21, 2024

JSON object

Loading...
9.8
Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 3.9 / Impact: 5.9
Source: NVD

Description

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When a message is sent to the HMI TCP socket, it is forwarded to the hmiProcessMsg function through the pendingQ, and may lead to remote code execution.

Affected (1)

Hmi 3 Control Panel Firmware
Configuration A
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
Before 7.2.5.7
Running on/withPlatform Versions
Swisslog Healthcare
Hmi 3 Control Panel
All versions