CVE-2020-12040
9.8
Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 3.9 / Impact: 5.9
Source: NVD
Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Affected (2)
Products: Baxter: Sigma Spectrum Infusion System Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| From 6.0 to 6.05 |
| Running on/with | Platform Versions |
|---|---|
Baxter Sigma Spectrum Infusion System | All versions |
References (2)
Source: ics-cert@hq.dhs.gov
Third Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource
Timeline
No history available yet.