CVE-2020-12035
4.9
Vector
CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L
Exploitability: 0.7 / Impact: 4.2
Source: NVD
Description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
Affected (2)
Products: Baxter: Prismaflex Firmware, Prismax Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Baxter Prismaflex | All versions |
Configuration B
| Vulnerable Software | Affected Versions |
|---|---|
| Before 3.0 |
| Running on/with | Platform Versions |
|---|---|
Baxter Prismax | All versions |
Related CWEs
References (3)
Source: af854a3a-2127-422b-91ae-364da2661108
Not Applicable
Timeline
No history available yet.