Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Affected (21)
Products: Medtronic: Mycarelink Monitor Firmware, Carelink Monitor Firmware, Carelink 2090 Firmware, Amplia Crt D Firmware, Claria Crt D Firmware, Compia Crt D Firmware, Concerto Crt D Firmware, Concerto Ii Crt D Firmware, Consulta Crt D Firmware, Evera Icd Firmware, Maximo Ii Crt D And Lcd Firmware, Mirro Icd Firmware, Nayamed Nd Icd Firmware, Primo Icd Firmware, Protecta Icd And Crt D Firmware, Secura Icd Firmware, Virtuoso Icd Firmware, Virtuoso Ii Icd Firmware, Visia Af Icd Firmware, Viva Crt D Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Version 24950 |
| Running on/with | Platform Versions |
|---|---|
Medtronic Mycarelink Monitor | All versions |
Configuration B
| Vulnerable Software | Affected Versions |
|---|---|
| Version 2490c |
| Running on/with | Platform Versions |
|---|---|
Medtronic Carelink Monitor | All versions |
Configuration C
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Carelink 2090 | All versions |
Configuration D
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Amplia Crt D | All versions |
Configuration E
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Claria Crt D | All versions |
Configuration F
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Compia Crt D | All versions |
Configuration G
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Concerto Crt D | All versions |
Configuration H
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Concerto Ii Crt D | All versions |
Configuration I
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Consulta Crt D | All versions |
Configuration J
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Evera Icd | All versions |
Configuration K
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Maximo Ii Crt D And Lcd | All versions |
Configuration L
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Mirro Icd | All versions |
Configuration M
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Nayamed Nd Icd | All versions |
Configuration N
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Primo Icd | All versions |
Configuration O
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Protecta Icd And Crt D | All versions |
Configuration P
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Secura Icd | All versions |
Configuration Q
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Virtuoso Icd | All versions |
Configuration R
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Virtuoso Ii Icd | All versions |
Configuration S
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Visia Af Icd | All versions |
Configuration T
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Medtronic Viva Crt D | All versions |
Related CWEs
CWE-284
Improper Access Control
The product does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
CWE-306
Missing Authentication for Critical Function
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.
CWE-862
Missing Authorization
The product does not perform an authorization check when an actor attempts to access a resource or perform an action.
References (4)
Source: ics-cert@hq.dhs.gov
MitigationThird Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryVDB Entry
Source: af854a3a-2127-422b-91ae-364da2661108
MitigationThird Party AdvisoryUS Government Resource