CVE-2019-18241
6.5
Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Exploitability: 2.8 / Impact: 3.6
Source: NVD
Description
In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub.
Affected (2)
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Philips Intellibridge Ec40 | All versions |
Configuration B
| Vulnerable Software | Affected Versions |
|---|---|
| All versions |
| Running on/with | Platform Versions |
|---|---|
Philips Intellibridge Ec80 | All versions |
References (2)
Source: ics-cert@hq.dhs.gov
Third Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource
Timeline