CVE-2019-13531
4.6
Vector
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
Exploitability: 0.9 / Impact: 3.6
Source: NVD
Description
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
Affected (3)
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Version 2.0.3 |
| Running on/with | Platform Versions |
|---|---|
Medtronic Valleylab Ft10 Energy Platform | All versions |
Configuration B
| Vulnerable Software | Affected Versions |
|---|---|
| Up to 1.20.2 |
| Running on/with | Platform Versions |
|---|---|
Medtronic Valleylab Ls10 Energy Platform | All versions |
References (3)
Source: ics-cert@hq.dhs.gov
Source: ics-cert@hq.dhs.gov
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource
Timeline
No history available yet.