CVE-2018-7510
9.8
Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 3.9 / Impact: 5.9
Source: NVD
Description
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.
Affected (1)
Products: Beaconmedaes: Scroll Medical Air Systems Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Before 4107600010.23 |
| Running on/with | Platform Versions |
|---|---|
Beaconmedaes Scroll Medical Air Systems | All versions |
Related CWEs
CWE-256
Plaintext Storage of a Password
Storing a password in plaintext may result in a system compromise.
CWE-522
Insufficiently Protected Credentials
The product transmits or stores authentication credentials, but it uses an insecure method that is susceptible to unauthorized interception and/or retrieval.