← Back

CVE-2018-14781

nvd nist
Published: Aug 13, 2018Modified: May 22, 2025

JSON object

Loading...
5.3
Vector
CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Exploitability: 1.6 / Impact: 3.6
Source: NVD

Description

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Affected (9)

508 Minimed Insulin Pump Firmware
522 Paradigm Real Time Firmware
722 Paradigm Real Time Firmware
523 Paradigm Revel Firmware
723 Paradigm Revel Firmware
523k Paradigm Revel Firmware
723k Paradigm Revel Firmware
551 Minimed 530g Firmware
751 Minimed 530g Firmware
Configuration A
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
508 Minimed Insulin Pump
All versions
Configuration B
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
522 Paradigm Real Time
All versions
Configuration C
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
722 Paradigm Real Time
All versions
Configuration D
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
523 Paradigm Revel
All versions
Configuration E
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
723 Paradigm Revel
All versions
Configuration F
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
523k Paradigm Revel
All versions
Configuration G
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
723k Paradigm Revel
All versions
Configuration H
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
551 Minimed 530g
All versions
Configuration I
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronicdiabetes
751 Minimed 530g
All versions

References (5)

Source: ics-cert@hq.dhs.gov
Third Party AdvisoryVDB Entry
Source: ics-cert@hq.dhs.gov
Third Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryVDB Entry
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryUS Government Resource

Timeline

No history available yet.