CVE-2018-10631

nvd nist
Published: Jul 13, 2018Modified: Aug 26, 2025

JSON object

Loading...
6.8
Vector
CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 0.9 / Impact: 5.9
Source: NVD

Description

The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.

Affected (2)

2 products
N'vision 8840 Firmware
N'vision 8870 Firmware
Configuration A
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronic
N'vision 8840
All versions
Configuration B
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
All versions
Running on/withPlatform Versions
Medtronic
N'vision 8870
All versions