An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
Affected (1)
Products: Abbott: Merlin@home Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Up to 8.0 |
| Running on/with | Platform Versions |
|---|---|
Abbott Merlin@home Ex1100 | All versions |
Abbott Merlin@home Ex1150 | All versions |