← Back

CVE-2017-5149

nvd nist
Published: Feb 13, 2017Modified: May 13, 2026

JSON object

Loading...
8.9
Vector
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H
Exploitability: 2.2 / Impact: 6.0
Source: NVD

Description

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

Affected (1)

1 product
Merlin@home Firmware
Configuration A
1 vulnerable · 2 platform
Vulnerable SoftwareAffected Versions
Up to 8.0
Running on/withPlatform Versions
Abbott
Merlin@home Ex1100
All versions
Abbott
Merlin@home Ex1150
All versions

References (4)

Source: ics-cert@hq.dhs.gov
Third Party AdvisoryVDB Entry
Source: ics-cert@hq.dhs.gov
MitigationThird Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryVDB Entry
Source: af854a3a-2127-422b-91ae-364da2661108
MitigationThird Party AdvisoryUS Government Resource

Timeline

No history available yet.