← Back

CVE-2017-11580

nvd nist
Published: Jul 2, 2019Modified: Nov 21, 2024

JSON object

Loading...
6.5
Vector
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Exploitability: 2.8 / Impact: 3.6
Source: NVD

Description

Blipcare Wifi blood pressure monitor BP700 10.1 devices allow memory corruption that results in Denial of Service. When connected to the "Blip" open wireless connection provided by the device, if a large string is sent as a part of the HTTP request in any part of the HTTP headers, the device could become completely unresponsive. Presumably this happens as the memory footprint provided to this device is very small. According to the specs from Rezolt, the Wi-Fi module only has 256k of memory. As a result, an incorrect string copy operation using either memcpy, strcpy, or any of their other variants could result in filling up the memory space allocated to the function executing and this would result in memory corruption. To test the theory, one can modify the demo application provided by the Cypress WICED SDK and introduce an incorrect "memcpy" operation and use the compiled application on the evaluation board provided by Cypress semiconductors with exactly the same Wi-Fi SOC. The results were identical where the device would completely stop responding to any of the ping or web requests.

Affected (1)

1 product
Configuration A
1 vulnerable · 1 platform
Vulnerable SoftwareAffected Versions
Up to bp700_10.1
Running on/withPlatform Versions
Blipcare
Wi Fi Blood Pressure Monitor
All versions

Related CWEs

References (6)

Source: cve@mitre.org
Mailing ListThird Party Advisory
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryVDB Entry
Source: af854a3a-2127-422b-91ae-364da2661108
ExploitThird Party Advisory
Source: af854a3a-2127-422b-91ae-364da2661108
Mailing ListThird Party Advisory

Timeline

No history available yet.