CVE-2017-11578
5.9
Vector
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
Exploitability: 2.2 / Impact: 3.6
Source: NVD
Description
It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.
Affected (1)
Products: Blipcare: Wi Fi Blood Pressure Monitor Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Up to bp700_10.1 |
| Running on/with | Platform Versions |
|---|---|
Blipcare Wi Fi Blood Pressure Monitor | All versions |
References (6)
http://packetstormsecurity.com/files/153225/Blipcare-Clear-Text-Communication-Memory-Corruption.html
Source: cve@mitre.org
Third Party AdvisoryVDB Entry
Source: cve@mitre.org
ExploitThird Party Advisory
http://packetstormsecurity.com/files/153225/Blipcare-Clear-Text-Communication-Memory-Corruption.html
Source: af854a3a-2127-422b-91ae-364da2661108
Third Party AdvisoryVDB Entry
Source: af854a3a-2127-422b-91ae-364da2661108
ExploitThird Party Advisory
Source: af854a3a-2127-422b-91ae-364da2661108
Mailing ListThird Party Advisory
Timeline
No history available yet.