Description
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Affected (1)
Products: Baxter: Sigma Spectrum Infusion System Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Version 6.05 |
| Running on/with | Platform Versions |
|---|---|
Baxter Sigma Spectrum Infusion System | All versions |
Baxter Wireless Battery Module | Version 16 |