CVE-2014-5432
9.8
Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploitability: 3.9 / Impact: 5.9
Source: NVD
Description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Affected (1)
Products: Baxter: Sigma Spectrum Infusion System Firmware
Configuration A
| Vulnerable Software | Affected Versions |
|---|---|
| Version 6.05 |
| Running on/with | Platform Versions |
|---|---|
Baxter Sigma Spectrum Infusion System | All versions |
Baxter Wireless Battery Module | Version 16 |
Related CWEs
CWE-287
Improper Authentication
When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.
CWE-592
DEPRECATED: Authentication Bypass Issues
This weakness has been deprecated because it covered redundant concepts already described in CWE-287.
References (2)
Source: ics-cert@hq.dhs.gov
MitigationThird Party AdvisoryUS Government Resource
Source: af854a3a-2127-422b-91ae-364da2661108
MitigationThird Party AdvisoryUS Government Resource
Timeline
No history available yet.